Mills B L, van de Voort F R, Usborne W R
J Assoc Off Anal Chem. 1983 Jul;66(4):1048-50. doi: 10.1016/s0315-5463(83)72185-1.
The Mojonnier method was compared with the conventional Soxhlet method for the determination of fat in 7 different meat products to assess its use as a standard reference method for calibration of commercial quantitative infrared transmission analyzers (e.g., Multispec M, Milkoscan 300 or 104). Results for the meat samples obtained by the Mojonnier method did not differ significantly from those obtained by the Soxhlet method. In addition, the Mojonnier method was less time-consuming and more precise than the Soxhlet; therefore, it can be used as a standard reference procedure for the calibration and assessment of infrared milk analyzers in their potential application to the rapid determination of fat in meat and meat products.
将莫乔尼尔法与传统索氏提取法用于测定7种不同肉类产品中的脂肪,以评估其作为商业定量红外透射分析仪(如Multispec M、Milkoscan 300或104)校准标准参考方法的适用性。通过莫乔尼尔法获得的肉类样品结果与通过索氏提取法获得的结果无显著差异。此外,莫乔尼尔法比索氏提取法耗时更少且更精确;因此,它可作为一种标准参考程序,用于校准和评估红外牛奶分析仪在快速测定肉类和肉类产品脂肪方面的潜在应用。