Bargo E S
J Assoc Off Anal Chem. 1983 Jul;66(4):864-6.
A reverse phase high pressure liquid chromatographic method for the determination of oxazepam in tablets and capsules was collaboratively studied by 9 laboratories. Collaborators were supplied with 6 samples that included synthetic and commercial formulations. Tablet and capsule composites are diluted with methanol and filtered. Oxazepam is determined at 254 nm by using a C18 column. Mean recoveries of oxazepam from synthetic tablet and capsule formulations were 97.2 and 99.0%, respectively. Mean coefficients of variation for tablets and capsules ranged from 1.85 to 2.86%. The method has been adopted official first action.
9个实验室合作研究了一种用于测定片剂和胶囊中奥沙西泮的反相高效液相色谱法。向合作人员提供了6个样品,包括合成制剂和市售制剂。片剂和胶囊的混合物用甲醇稀释并过滤。使用C18柱在254nm波长处测定奥沙西泮。奥沙西泮在合成片剂和胶囊制剂中的平均回收率分别为97.2%和99.0%。片剂和胶囊的平均变异系数在1.85%至2.86%之间。该方法已被官方首次采用。