Rejger V, Burm A G, Spierdijk J, Strokap H
Br J Anaesth. 1981 Dec;53(12):1359-63. doi: 10.1093/bja/53.12.1359.
A scavenging system which can be attached to a variety of anaesthetic machines and ventilators is described. Gas venting can occur via a central vacuum system, a suitable injector, fan or pump, through the exhaust duct of an air-conditioning system, or by a passive system. The resistance and back pressures applied to the anaesthetic apparatus are negligible. Leakage tests showed that the collected gases and vapours were substantially removed by the scavenging device.
描述了一种可连接到各种麻醉机和呼吸机的清除系统。气体排放可通过中央真空系统、合适的注射器、风扇或泵,通过空调系统的排气管,或通过被动系统进行。施加在麻醉设备上的阻力和背压可忽略不计。泄漏测试表明,清除装置可基本清除收集到的气体和蒸汽。