Huang S C, Hoffman E J, Phelps M E, Kuhl D E
J Comput Assist Tomogr. 1980 Dec;4(6):819-26. doi: 10.1097/00004728-198012000-00014.
Effect of insufficient transverse sampling on quantitative positron emission computed tomography (ECT) was investigated with computer simulation and measurements on parallel bar and line source phantoms. Aliasing artifacts were observed and were found to be dependent on both the configuration and the location of imaged objects. Images of parallel bar phantoms were found to have aliasing artifacts similar in characteristics to aliasing on one-dimensional signals. In line source images, aliasing effects were manifested as variations in amplitude and full width at half maximum resolution (FWHM) for sources at even slightly different locations in the field of view. It was found that employing sampling distances smaller than one-third of the intrinsic detector FWHM eliminated noticeable aliasing artifacts. Image resolution was also found to be affected by the sampling distance. For a sampling distance equal to one-half of the intrinsic detector FWHM, the imaging FWHM is about 10% worse than the intrinsic FWHM. Selection of sampling distance in noisy environments is discussed. Parallel bar phantoms are shown to have advantages over line sources in the evaluation of sampling and resolution performance of ECT scanners.
通过计算机模拟以及对平行棒和线源模型的测量,研究了横向采样不足对定量正电子发射计算机断层扫描(ECT)的影响。观察到了混叠伪影,发现其取决于成像物体的配置和位置。发现平行棒模型的图像具有与一维信号上的混叠特征相似的混叠伪影。在线源图像中,对于视野中即使位置稍有不同的源,混叠效应表现为幅度变化和半高宽分辨率(FWHM)变化。发现采用小于探测器固有FWHM三分之一的采样距离可消除明显的混叠伪影。还发现图像分辨率受采样距离影响。对于等于探测器固有FWHM一半的采样距离,成像FWHM比固有FWHM约差10%。讨论了在噪声环境中采样距离的选择。在评估ECT扫描仪的采样和分辨率性能方面,平行棒模型显示出优于线源的优势。