A boundary method for attenuation correction in positron computed tomography.

作者信息

Huang S C, Carson R E, Phelps M E, Hoffman E J, Schelbert H R, Kuhl D E

出版信息

J Nucl Med. 1981 Jul;22(7):627-37.

PMID:6973010
Abstract

A new method for attenuation correction in positron computed tomography (PCT) has been developed, and it can improve the quality of PCT images. The method requires a short transmission scan by the PCT system. Then boundaries between tissues with significantly different attenuation coefficients are determined from the transmission image by edge-finding techniques. Attenuation correction factors(ACF) are then calculated using these boundaries and the average attenuation coefficients within the enclosed regions. The method has been tested on computer-simulated data, on scans of phantoms, and on patient studies, and has been found effective in reducing the random noise in transmission measurements and in providing more accurate ACFs than the method using geometric attenuation correction. As a result, transmission scan times can be shortened, inconvenience to patients is reduced, and PCT images are improved.

摘要

已经开发出一种用于正电子计算机断层扫描(PCT)中衰减校正的新方法,它可以提高PCT图像的质量。该方法需要PCT系统进行一次短时间的透射扫描。然后通过边缘检测技术从透射图像中确定具有显著不同衰减系数的组织之间的边界。接着利用这些边界和封闭区域内的平均衰减系数来计算衰减校正因子(ACF)。该方法已经在计算机模拟数据、体模扫描和患者研究中进行了测试,并且已发现它在降低透射测量中的随机噪声以及提供比使用几何衰减校正的方法更准确的ACF方面是有效的。结果,可以缩短透射扫描时间,减少给患者带来的不便,并且改善PCT图像。