在正电子发射断层扫描中使用计数受限传输数据进行衰减校正。

Attenuation correction using count-limited transmission data in positron emission tomography.

作者信息

Meikle S R, Dahlbom M, Cherry S R

机构信息

Department of Nuclear Medicine, Royal Prince Alfred Hospital, Sydney, Australia.

出版信息

J Nucl Med. 1993 Jan;34(1):143-50.

PMID:8418258
Abstract

Poisson noise in transmission data can have a significant influence on the statistical uncertainty of PET measurements, particularly at low transmission count rates. In this paper, we investigate the effect of transmission data processing on noise and quantitative accuracy of reconstructed PET images. Differences in spatial resolution between emission and transmission measurements due to transmission data smoothing are shown to have a significant influence on quantitative accuracy and can lead to artifacts in the reconstructed image. In addition, the noise suppression of this technique is insufficient to greatly reduce transmission scan times. Based on these findings, improved strategies for processing count-limited transmission data have been developed, including a method using segmentation of attenuation images. Using this method, accurate attenuation correction can be performed using transmission scan times as low as 2 min without increasing noise in reconstructed PET images.

摘要

传输数据中的泊松噪声会对PET测量的统计不确定性产生重大影响,尤其是在低传输计数率时。在本文中,我们研究了传输数据处理对重建PET图像的噪声和定量准确性的影响。由于传输数据平滑,发射和传输测量之间的空间分辨率差异对定量准确性有重大影响,并可能导致重建图像中出现伪影。此外,该技术的噪声抑制不足以大幅减少传输扫描时间。基于这些发现,已开发出改进的处理计数受限传输数据的策略,包括一种使用衰减图像分割的方法。使用这种方法,可以在低至2分钟的传输扫描时间下进行准确的衰减校正,而不会增加重建PET图像中的噪声。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验