Link J, Voll K
Anaesthesist. 1981 Apr;30(4):210-2.
The inspiratory resistance, which a patient has to overcome while doing SIMV and CPAP with a Servoventilator 900 B, can be lowered by a simple modification of the ventilator. This modification is done by measuring the ventilation-pressure at the patients side of the humidifier. Measurements of pressure and flow demonstrated that, with a flow of 11 s-1 a 66%-reduction of the inspiratory resistance can be achieved.