A new computerized fluoroscopy technique for isolating low image contrast, which results during the flow of an intravenously administered bolus of iodinated contrast materials, is proposed. The technique requires the application of one of a family of imaging algorithms which have been designed to isolate time varying image contrast. This family of imaging algorithm is described, as is the way in which each isolates a particular range of temporal frequency components associated with the bolus's flow through various cardiovascular structures. The implementation of these algorithms in real time using appropriate digital recursive filtering techniques is described. The architecture of a dedicated hardwired computerized fluoroscopy apparatus which would incorporate these algorithms is also presented.
本文提出了一种新的计算机荧光透视技术,用于分离静脉注射碘造影剂团注过程中产生的低图像对比度。该技术需要应用一系列成像算法中的一种,这些算法旨在分离随时间变化的图像对比度。本文描述了这一系列成像算法,以及每种算法分离与造影剂团注通过各种心血管结构相关的特定时间频率分量范围的方式。还描述了如何使用适当的数字递归滤波技术实时实现这些算法。本文还介绍了一种将包含这些算法的专用硬连线计算机荧光透视设备的架构。