Demacker P N, Hijmans A G, van Sommeren-Zondag D F, Jansen A P
Clin Chem. 1982 Jan;28(1):155-7.
Liquid serum pools with low-, normal-, or above-normal concentration of high-density lipoprotein cholesterol were prepared by selection and dilution of sera having low lipoprotein content or by enrichment with concentrated high-density lipoprotein. Stability of the serum pools depended on storage conditions. On storage between -2 and -12 degrees C, the high-density lipoprotein cholesterol content decreased. At a constant temperature of -19 to -21 degrees C the concentrations remained stable for more than 18 months. The imprecision (CV) of the high-density lipoprotein cholesterol assay during this period as established with enzymic cholesterol analysis (Clin. Chem. 26: 1780-1786, 1980) of these serum pools was between 2.7 and 4.8% (n = 51). Serum pools prepared and stored as described are suitable for internal quality-control procedures. In external quality-control trials these sera may be superior to the commercially available lyophilized lipid control sera.
通过选择和稀释低脂蛋白含量的血清或用浓缩高密度脂蛋白进行富集,制备了高密度脂蛋白胆固醇浓度低、正常或高于正常的液体血清池。血清池的稳定性取决于储存条件。在-2至-12摄氏度之间储存时,高密度脂蛋白胆固醇含量会降低。在-19至-21摄氏度的恒定温度下,浓度可保持稳定超过18个月。在此期间,通过对这些血清池进行酶促胆固醇分析(临床化学26:1780-1786,1980)确定的高密度脂蛋白胆固醇测定的不精密度(CV)在2.7%至4.8%之间(n = 51)。按所述方法制备和储存的血清池适用于内部质量控制程序。在外部质量控制试验中,这些血清可能优于市售的冻干脂质对照血清。