Immunoinhibition and automated column chromatography compared for assay of creatine kinase isoenzyme MB in serum.

作者信息

Bayer P M, Boehm M, Hajdusich P, Hotschek H, Koehn H, Unger W, Wider G

出版信息

Clin Chem. 1982 Jan;28(1):166-9.

PMID:7055904
Abstract

We examined sera from six different groups of patients for CK-MB activity by means of two commercially available tests, an immunoinhibition method (E. Merck) and the CK-MB test as used with the aca (Du Pont). In the first group of patients (suspicion of myocardial infarction) the correlation between the two methods was good: r = 0.9191, y = 1.068x -- 0.888, x = 18.7 U/L, y = 19.0 U/L. In the second group, patients with high adenylate kinase activity, no interference was detectable on the aca, whereas the immunoinhibition method yielded falsely high CK-MB values. The third group consisted of persons with macro-CK-BB in their serum. In the immunoinhibition test these patients usually showed a high CK-MB:total CK ratio, whereas such results were rarely found for the aca. The fourth group, patients with a different electrophoretic mobility of their CK-isoenzymes (migration of an active band towards the cathode), were detected by the immunoinhibition method (high ratio of CK-MB to total CK), but not with the aca. In the presence of free CK-BB (group five) the immunoinhibition test resulted in "falsely" high CK-MB values, whereas CK-BB was retained on the column of the aca. In skeletal muscle diseases (group six) results by the two methods differed, values for CK-MB on the aca being much higher. It was demonstrated experimentally that this was due to CK-MM with altered surface charge.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验