Giles A F
Clin Endocrinol (Oxf). 1982 Jan;16(1):101-5. doi: 10.1111/j.1365-2265.1982.tb03153.x.
A convenient, sensitive radioimmunoassay of free thyroxine in serum dialysates is described. The method utilizes a solid phase separation system (pre-formed double antibody) and a relatively short incubation period (220 min) with a staggered addition of tracer. Blanks were low and consistent. The normal range for non-pregnant euthyroid samples (n=59) was 11-23 pmol/l. Third trimester pregnancy samples were mostly within the normal range but at the lower end. Patients on T4 replacement showed a much wider variation in free T4 levels with many samples in the hyperthyroid region. Some hypothyroid samples had undetectable free T4 levels and hyperthyroid samples were frequently greater than 80 pmol/l.
本文描述了一种用于血清透析液中游离甲状腺素的简便、灵敏的放射免疫测定法。该方法采用固相分离系统(预形成的双抗体),并通过错时添加示踪剂实现相对较短的温育期(220分钟)。空白值较低且稳定。非妊娠甲状腺功能正常样本(n = 59)的正常范围为11 - 23 pmol/l。妊娠晚期样本大多在正常范围内,但处于较低水平。接受T4替代治疗的患者游离T4水平变化范围更大,许多样本处于甲状腺功能亢进区域。一些甲状腺功能减退样本的游离T4水平无法检测到,而甲状腺功能亢进样本通常大于80 pmol/l。