Wu V C, Nickell W T, Bhagat P K
Aviat Space Environ Med. 1982 Apr;53(4):375-8.
An ultrasonic plethysmograph, which gives improved performance over the standard Whitney Strain Gauge, is described. This instrument monitors dimension changes in human limbs by measuring the transit times of acoustic pulses across two chords of the limb. In the case of a small uniform expansion, the percentage change in limb volume is shown to be proportional to twice the percentage change in either of the measured chords. Measurement of two chords allows correction for possible non-uniform expansion. In addition, measurement of two chords allows an estimate of the absolute cross-sectional area of the limb. The developed instrument incorporates a microprocessor, which performs necessary calculation and control functions. Use of the microprocessor allows the instrument to be self-calibrating. In addition, the device can be easily reprogrammed to incorporate improvements in operating features or computational schemes.