与美国药典溶出度试验II装置2相关的系统误差:溶出杯曲率与球体曲率偏差的影响。

Systematic error associated with apparatus 2 of the USP dissolution test II: Effects of deviations in vessel curvature from that of a sphere.

作者信息

Cox D C, Wells C E, Furman W B, Savage T S, King A C

出版信息

J Pharm Sci. 1982 Apr;71(4):395-9. doi: 10.1002/jps.2600710405.

Abstract

Dissolution vessels made from glass or plastic are recognized by the USP as being suitable for dissolution testing. Glass vessels with a bottom inside curvature flatter than that of a sphere can cause a high bias in dissolution results; vessels with a steeper curvature can cause a low bias. The inside bottom curvature of plastic vessels adhered closely to the curvature of a sphere. The plastic vessels are preferable for use if the drug is not adsorbed and the vessel is not attacked by the dissolution medium. Bias in results between individual positions of a dissolution apparatus was traced to two shafts which were not vertical.

摘要

美国药典认可由玻璃或塑料制成的溶出度测定容器适用于溶出度测试。底部内曲率比球体更平缓的玻璃容器会导致溶出度结果出现较大偏差;曲率更陡的容器会导致偏差较小。塑料容器的底部内曲率与球体曲率紧密贴合。如果药物不会被吸附且溶出介质不会侵蚀容器,则使用塑料容器更为可取。溶出度测定仪各个位置之间的结果偏差可追溯到两根不垂直的轴。