用光纤激光多普勒血流仪测量冠状窦血流量。

Measurement of coronary sinus blood flow by fiber-optic laser Doppler anemometry.

作者信息

Kilpatrick D, Linderer T, Sievers R E, Tyberg J V

出版信息

Am J Physiol. 1982 Jun;242(6):H1111-4. doi: 10.1152/ajpheart.1982.242.6.H1111.

Abstract

The Doppler shift of a laser has been used to measure the velocity of blood in the coronary sinus of the dog. The laser was delivered along a fine fiber-optic probe, which was also used to receive the reflected light. The Doppler shift was measured from the output of a digital spectrum analyzer. This measurement, proportional to the velocity of blood, was correlated with the blood flow measured by an electromagnetic flowmeter. Studies were performed on 112 successful maneuvers in eight consecutive anesthetized dogs. In the minimally damped electromagnetic flow traces, the correlation between coronary sinus blood flow velocity and coronary sinus blood flow was high (r greater than 0.95). This measurement will give absolute flow values only if the exact dimensions of the coronary sinus are known. The linear relationship of coronary sinus velocity and flow supports the postulate that coronary sinus dimension remains similar during normal and high flows. This technique, fiber-optic laser Doppler anemometry, is an accurate measure of relative coronary sinus blood flow. It is suitable for use in human coronary sinuses, especially to monitor the effect of interventions.

摘要

激光的多普勒频移已被用于测量狗冠状窦内的血流速度。激光通过一根精细的光纤探头传输,该探头也用于接收反射光。多普勒频移是从数字频谱分析仪的输出中测量的。这一与血流速度成正比的测量值与用电磁流量计测得的血流量相关。对八只连续麻醉的狗进行了112次成功的操作研究。在最小阻尼的电磁血流记录中,冠状窦血流速度与冠状窦血流量之间的相关性很高(r大于0.95)。仅当冠状窦的确切尺寸已知时,这种测量才能给出绝对流量值。冠状窦速度与流量的线性关系支持了这样的假设,即在正常流量和高流量期间冠状窦尺寸保持相似。这种技术,即光纤激光多普勒风速测定法,是相对冠状窦血流量的一种准确测量方法。它适用于人体冠状窦,尤其适用于监测干预效果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验