Suppr超能文献

A "one-minute" direct bilirubin determination using the Abbott ABA-100 Bichromatic Analyser.

作者信息

Chu S Y, Kaczmareck A, Turkington V E

出版信息

Clin Biochem. 1982 Apr;15(2):113-5. doi: 10.1016/s0009-9120(82)96283-x.

Abstract

We describe a "one-minute" direct bilirubin procedure using the ABA-100. The procedure correlates well with the manual Jendrassik-Grof method and gives good precision at a relatively low direct bilirubin level. In comparison with certain other Abbott ABA-100 direct bilirubin methods, the proposed procedure appears to be the least affected by the presence of high unconjugated bilirubin levels.