Evaluation of a miniature CdTe detector for monitoring left ventricular function.

作者信息

Harrison K S, Liu X, Han S T, Camargo E E, Wagner H N

出版信息

Eur J Nucl Med. 1982;7(5):204-6. doi: 10.1007/BF00256464.

Abstract

A miniature CdTe probe interfaced to a microcomputer was used to measure left ventricular ejection fraction (LVEF) in 25 patients. LVEF obtained with the CdTe module, in the beat-to-beat mode, or the integrated gated mode agreed well with LVEF obtained with a gamma camera (r = 0.80; r = 0.82 respectively). Similarly, LVEF by CdTe probe agreed with LVEF obtained by gated equilibrium studies performed with a computerized NaI probe. The CdTe probe can provide comparable measurement of LVEF at a fraction of the cost of a camera-computer system and, being small and lightweight, the CdTe probe is adaptable for monitoring patients in intensive care facilities.