Suppr超能文献

使用间断流单针系统对不同AN 69膜装置进行血液滤过的评估。

Evaluation of hemofiltration with different AN 69 membrane devices using a discontinuous flow-single needle system.

作者信息

De Paepe M, Ringoir S

出版信息

Int J Artif Organs. 1982 Mar;5(2):87-91.

PMID:7095885
Abstract

Unipuncture hemofiltration using the double headed pump (Bellco BL 760) was performed in a patient who acted as her own control for 10 sessions with a 1.0 m2 AN69 membrane (group A), 10 sessions with a 1.4 m2 AN69 membrane (group B) and 10 sessions with a 1.0 m2 AN69 membrane Filtral (group C). A two needle standard hemofiltration technique (Hemoprocessor Sartorius) was used in this patient for 10 sessions with a 1.0 m2 AN69 membrane (group D). To obtain the same amount of ultrafiltration fluid the filtration time was respectively 266, 196, 214 and 260 minutes in groups A, B, C and D.

摘要

使用双头泵(Bellco BL 760)对一名患者进行单针血液滤过,该患者作为自身对照,使用1.0 m²的AN69膜进行10次治疗(A组),使用1.4 m²的AN69膜进行10次治疗(B组),使用1.0 m²的AN69膜Filtral进行10次治疗(C组)。该患者使用两针标准血液滤过技术(Hemoprocessor Sartorius),使用1.0 m²的AN69膜进行10次治疗(D组)。为获得相同量的超滤液,A、B、C和D组的滤过时间分别为266、196、214和260分钟。