Schmidt W F
J Assoc Off Anal Chem. 1982 May;65(3):753-5.
A high performance liquid chromatographic procedure has been developed for the assay of phenprocoumon in tablets. In comparison to the present official USP assay procedure, it is equivalent in precision and accuracy and is faster and more specific. A mobile phase consisting of a 1% solution of acetic acid in acetonitrile-water (4 + 3) separates phenprocoumon from warfarin internal standard on a 6 micrometer octadecylsilane (ODS) column with UV detection at 311 nm. The method enables the concurrent determination of phenprocoumon and possible contaminants such as salicylic acid.
已开发出一种高效液相色谱法用于测定片剂中的苯丙香豆素。与现行美国药典法定测定方法相比,该方法在精密度和准确度方面相当,且速度更快、特异性更强。流动相由1%的乙酸乙腈 - 水(4 + 3)溶液组成,在6微米十八烷基硅烷(ODS)柱上,于311 nm处进行紫外检测,可将苯丙香豆素与华法林内标分离。该方法能够同时测定苯丙香豆素以及可能的污染物,如水杨酸。