Hamlin C, Ackerman E
Clin Chem. 1978 Nov;24(11):2013-7.
Prospective analysis (by use of commercially available kits) of 339 validated admission specimens from patients consecutively admitted to a cardiac monitoring unit revealed sensitivities of 68 and 63% for two electrophoretic assays (Bioware and Helena) and up to 73 and 77% for two chromatographic assays (Roche and Worthington). Specificities were 94.8, 96,6, 90.2, and 89.7% and efficiencies were 90.2, 90.7, 87.0, and 87.0, respectively. Although these results differ from early reports, they agree well with several recent studies. However, 86 (25%) of the specimens gave at least one discrepant result based on clinical evaluation of the patients, and 68 (20%) gave at least one discrepant result between methods.