Baughman D J, Lytwyn A
Thromb Res. 1982 Apr 1;26(1):1-12. doi: 10.1016/0049-3848(82)90145-1.
Some properties of a one-stage, chromogenic clotting assay using tissue thromboplastin as an activator are described. The chromogenic assay uses S-2238 as substrate and estimates thrombin by the rate of absorbance produced. Under assay conditions, it was observed that the log (thrombin or absorbance/min) is linear with time. The linearity appeared to extend from zero thrombin up to maximum thrombin concentration for all concentrations of each plasma type examined including factor V and factor VII deficient plasmas. Since the linear slope, b, appears to be a measure of the rate of thrombin production, it was called the Thrombin Activation Rate Constant (TARC). For dilutions of normal plasma log (b) appears to be linearly related to log (plasma concentration) and to log (PT) where PT is the standard one stage clotting time. For plasmas deficient in extrinsic clotting factors, b was linearly related to log (factor concentrations). The chromogenic assay was most sensitive to changes in prothrombin; least sensitive to changes in factor VII; and unaffected by changes in concentration of factors VIII and IX. The standard PT was least sensitive to prothrombin and more sensitive to factors X and VII. For oral anticoagulated plasmas log b and log PT were linearly correlated with a regression slope which was more negative than that from normal plasma. From theses results it was concluded that TARC is a chromogenic assay sensitive to all factors in the extrinsic pathway; that TARC might be used as both a screening assay for extrinsic deficiencies and a monitoring assay for anticoagulants; and that TARC may be more sensitive than the standard PT to the defects from oral anticoagulation.
描述了一种使用组织凝血活酶作为激活剂的一步法发色凝血测定的一些特性。该发色测定使用S-2238作为底物,并通过产生的吸光度速率来估计凝血酶。在测定条件下,观察到log(凝血酶或吸光度/分钟)与时间呈线性关系。对于所检测的每种血浆类型的所有浓度,包括因子V和因子VII缺乏血浆,线性似乎从零凝血酶延伸至最大凝血酶浓度。由于线性斜率b似乎是凝血酶产生速率的一种度量,因此将其称为凝血酶激活速率常数(TARC)。对于正常血浆的稀释液,log(b)似乎与log(血浆浓度)以及log(PT)呈线性相关,其中PT是标准的一步凝血时间。对于外源性凝血因子缺乏的血浆,b与log(因子浓度)呈线性相关。发色测定对凝血酶原的变化最敏感;对因子VII的变化最不敏感;并且不受因子VIII和IX浓度变化的影响。标准PT对凝血酶原最不敏感,对因子X和VII更敏感。对于口服抗凝血浆,log b和log PT呈线性相关,其回归斜率比正常血浆的更负。从这些结果得出结论,TARC是一种对外源性途径中的所有因子敏感的发色测定;TARC可既用作外源性缺乏的筛查测定,又用作抗凝剂的监测测定;并且TARC可能比标准PT对口服抗凝引起的缺陷更敏感。