A system has been built to allow up to four scintillation cameras to be connected in parallel with two computers, yet provide for selection of individual pairs of camera/computer combinations with the exclusion of unwanted connections. Since the distances involved were relatively long, signal drivers were used to transmit the analog signals in a differential mode over multiple twin-paired lines rather than coaxial cable. By transmitting the signals differentially, any noise induced on the signal was subtracted at the receiver. The principal features of this system, which has been installed in two institutions, are described. A problem that occurred in connection with one camera is presented (also its solution), with a suggestion concerning parameters to be assessed at the time of system acceptance.
已经构建了一个系统,该系统允许最多四台闪烁相机与两台计算机并行连接,但要能够选择单独的相机/计算机组合对,同时排除不必要的连接。由于涉及的距离相对较长,因此使用信号驱动器以差分模式通过多条双绞线路传输模拟信号,而不是使用同轴电缆。通过差分传输信号,在接收器处可以减去信号上感应的任何噪声。本文描述了已安装在两个机构中的该系统的主要特征。还介绍了与一台相机相关出现的一个问题(以及其解决方案),并针对系统验收时要评估的参数提出了建议。