Suppr超能文献

一种用于使用安格相机逐搏测量心脏功能的双准直器设计。

A dual collimator design for beat-to-beat measurement of cardiac performance with an Anger camera.

作者信息

Heller S L, Goodwin P N, Scharf S C, Hardoff R, Blaufox M D

出版信息

J Nucl Med. 1982 Sep;23(9):836-9.

PMID:7108631
Abstract

A dual collimator was designed for an Anger camera to permit measurement of cardiac performance on a beat-to-beat basis. Special all-purpose (SAP) and special high-sensitivity (SHS) collimator sections can be interchanged without movement of the patient. Thus, left-ventricular regions of interest delineated on SAP multigated images can be transferred to SHS dynamic images to generate beat-to-beat volume curves. Preliminary balloon studies demonstrated an excellent correlation between ejection fractions calculated with the two collimators: r greater than 0.99, n = 17, p less than 0.001. Varying the volume of an adjacent "right ventricle" balloon failed to alter significantly the count rate from the "left ventricle" balloon's region of interest. Preliminary results on 12 patients, comparing standard-camera ejection fractions with average beat-to-beat ejection fractions, showed that is is possible to measure cardiac function on a beat-to-beat basis with a single-crystal gamma camera. There was minimal difference between the ejection fractions calculated by the dual-collimator method and a standard gated technique (r = 0.98, n = 12, p less than 0.001).

摘要

为一台安格相机设计了一种双准直器,以便能够逐搏测量心脏功能。特殊通用(SAP)准直器部分和特殊高灵敏度(SHS)准直器部分可以在不移动患者的情况下互换。因此,在SAP多门控图像上划定的左心室感兴趣区可以转移到SHS动态图像上,以生成逐搏容积曲线。初步的球囊研究表明,用这两种准直器计算的射血分数之间具有极佳的相关性:r大于0.99,n = 17,p小于0.001。改变相邻“右心室”球囊的容积,未能显著改变来自“左心室”球囊感兴趣区的计数率。对12名患者进行的初步研究结果,将标准相机射血分数与平均逐搏射血分数进行比较,结果表明,使用单晶γ相机能够逐搏测量心脏功能。双准直器法与标准门控技术计算的射血分数之间差异极小(r = 0.98,n = 12,p小于0.001)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验