Suppr超能文献

人工心脏瓣膜的体外评估:异常与局限性

In vitro evaluation of prosthetic heart valves: anomalies and limitations.

作者信息

Tindale W B, Black M M, Martin T R

出版信息

Clin Phys Physiol Meas. 1982 May;3(2):115-30. doi: 10.1088/0143-0815/3/2/002.

Abstract

The in vitro evaluation of fifteen commercially available prosthetic heart valves is described. The group comprised both mechanical and bioprosthetic valves. The experimental studies were carried out in a steady flow circuit which included a flexible valve test section, the geometry of which was modelled on the aortic anatomy. The fluidic performance of the valves was examined over a range of flows and in particular, the phenomenon of "pressure recovery" distal to the valve was studied in depth. A number of theoretical parameters which might be used to define valve performance were evaluated but none was found to be ideal. The conclusions of the study emphasise the limitations of such in vitro testing and the danger of extrapolating such experimental data to the possible valve performance in vivo.

摘要

本文描述了对15种市售人工心脏瓣膜的体外评估。该组包括机械瓣膜和生物瓣膜。实验研究在一个稳流回路中进行,该回路包括一个柔性瓣膜测试段,其几何形状以主动脉解剖结构为模型。在一系列流量条件下检查了瓣膜的流体性能,特别是深入研究了瓣膜远端的“压力恢复”现象。评估了一些可能用于定义瓣膜性能的理论参数,但未发现理想的参数。该研究的结论强调了这种体外测试的局限性,以及将此类实验数据外推至体内瓣膜可能性能的危险性。