England J M, Chetty M C, Chadwick R, Woodhead G B
Clin Lab Haematol. 1982;4(2):187-99.
The Ortho ELT-8 is an automated blood counter which appears to be safe, precise and free from carry-over. Red and white cell results generally agree with those on the Coulter Counter, Model S, though discrepancies were noted with the WBC and PCV. Reference methods showed the Model S WBC results tended to be inaccurate on the discrepant samples though neither instrument was predominantly responsible for the PCV discrepancies. The ELT-8 platelet count tended to be higher than with the Thrombocounter/Thrombofuge system. When packed cells were diluted in autologous plasma serious variations in the red cell indices (MCV, MCH & MCHC) were first found due to incorrect voltage-frequency converter setting. Even after this setting had been corrected some variations in the MCH and MCHC were still apparent. In the UK National External Quality Assessment Scheme the ELT-8 results on animal bloods did not agree with those produced by Model S users; this discrepancy probably being due to differences between the light-scattering and aperture-impedance technology.
Ortho ELT - 8是一种自动血液计数器,似乎安全、精确且无残留问题。红细胞和白细胞计数结果通常与库尔特S型计数器的结果一致,不过白细胞和红细胞压积出现了差异。参考方法显示,在存在差异的样本上,S型计数器的白细胞计数结果往往不准确,不过两种仪器都不是导致红细胞压积差异的主要原因。ELT - 8的血小板计数往往高于血栓计数器/血栓离心机系统。当将压积细胞在自体血浆中稀释时,由于电压 - 频率转换器设置错误,首次发现红细胞指数(平均红细胞体积、平均红细胞血红蛋白含量和平均红细胞血红蛋白浓度)出现严重变化。即使在该设置校正后,平均红细胞血红蛋白含量和平均红细胞血红蛋白浓度仍有一些变化。在英国国家外部质量评估计划中,ELT - 8对动物血液的检测结果与S型计数器用户的结果不一致;这种差异可能是由于光散射技术和小孔阻抗技术之间的差异所致。