Stereospecific lactate absorption during peritoneal dialysis.

作者信息

Rubin J, Adair C, Johnson B, Bower J D

出版信息

Nephron. 1982;31(3):224-8. doi: 10.1159/000182651.

Abstract

Patients undergoing peritoneal dialysis were studied to determine if peritoneal absorption was selective. Dialysis was performed using dialysate exchange schedules similar to those for intermittent peritoneal dialysis and continuous ambulatory peritoneal dialysis. The clearance rate from the peritoneal cavity during hourly dialysate exchanges was 6.2 ml/min for D (-)-lactate and 8.7 ml/min for L(+)-lactate (p less than 0.01). L(+)-Lactate disappeared more rapidly from the dialysate during the long-cycle exchanges. Our results suggest that clearance of lactate from the peritoneal cavity is relatively stereospecific and raises the question of selective absorption for other organic anions.

摘要

对接受腹膜透析的患者进行了研究,以确定腹膜吸收是否具有选择性。使用与间歇性腹膜透析和持续性非卧床腹膜透析相似的透析液交换方案进行透析。在每小时的透析液交换过程中,腹膜腔对D(-)-乳酸的清除率为6.2毫升/分钟,对L(+)-乳酸的清除率为8.7毫升/分钟(p小于0.01)。在长周期交换过程中,L(+)-乳酸从透析液中消失得更快。我们的结果表明,腹膜腔对乳酸的清除具有相对的立体特异性,并引发了关于其他有机阴离子选择性吸收的问题。