Smith J, Svenjak D, Turrell J, Vlastelica D
Clin Chem. 1982 Sep;28(9):1867-72.
A major instrumental limitation to rapid, "random-access" clinical analysis has been cross-contamination at sampling and reagent probes. Use of an immiscible, nonreactive fluid as a positive barrier between the liquid sample and reagent and the interior and exterior surfaces of their respective probes provides an inert, deformable surface that both prevents carryover and ensures accurate delivery. Being totally nonreactive under clinical chemistry conditions, the fluid does not affect the ensuing analysis. Application of this technology to the Technicon RA-1000 (trademark of Technicon Instruments Corp.) system allows transfer of a liquid sample and any of 14 liquid reagents to the reaction vessel with a carryover of less than 1/10(6) with a precision (CV) of 0.5%. Its use provides for a rapid analysis rate of 240 tests per hour, and for transfer of microliter quantities of serum, with carryover of less than 1/8000, again at an average CV of 0.5%.
快速“随机存取”临床分析的一个主要仪器限制因素是采样和试剂探针处的交叉污染。使用一种不混溶、无反应性的流体作为液体样品与试剂以及它们各自探针的内、外表面之间的正向屏障,可提供一个惰性、可变形的表面,既能防止残留,又能确保准确加样。该流体在临床化学条件下完全无反应性,不会影响后续分析。将此技术应用于Technicon RA - 1000(Technicon仪器公司的商标)系统,可将液体样品和14种液体试剂中的任何一种转移至反应容器,残留率小于1/10⁶,精密度(CV)为0.5%。使用该技术可实现每小时240次检测的快速分析速度,并且在转移微升量血清时,残留率小于1/8000,平均CV同样为0.5%。