Lua A C, Chu D K, Vlastelica D
Miles Inc., Diagnostic Division, Tarrytown, NY 10591.
Ther Drug Monit. 1994 Oct;16(5):495-8. doi: 10.1097/00007691-199410000-00009.
We report the determination of digoxin concentration in serum with Microgenics Cedia digoxin reagents on the Technicon CHEM 1. The Technicon CHEM 1 clinical chemistry analyzer has a throughput of 720 tests per hour and uses only 7 microliters each of two reagents. A 100 test kit can perform 2,640 tests. The within-run coefficient of variation (CV) range is 2.3-0.9% and the total CV is 6.3-2.9% at concentrations tested ranging from 1.10 to 2.94 ng/ml. The results of the Technicon CHEM 1 (y) assay correlated well with those by the Technicon RA 1000 system (x) with 31 clinical serum samples (y = -0.03 + 1.11x, r = 0.96). We concluded that the Cedia digoxin assay on the Technicon CHEM 1 provides a very cost-effective, precise, rapid, and accurate means to determine digoxin concentration in serum.