Suppr超能文献

子宫形态与宫内节育器设计

Uterine geometry and IUCD design.

作者信息

Hasson H M

出版信息

Br J Obstet Gynaecol. 1982 Sep;89(Suppl 4):1-10. doi: 10.1111/j.1471-0528.1982.tb15064.x.

Abstract

The endometrial cavity is an isosceles trapezoid that changes in shape with the phase of the cycle. The distance from the fundus to where the cavity narrows to 12 mm is the effective length. This and the total length of the uterine cavity can be measured with the Wing Sound. IUCD complications are more frequent if the length of a device exceeds the effective length or the width of the uterine cavity. Devices must be designed to fit the uterine cavity and to conform to physiological changes in the shape of the organ.

摘要

子宫内膜腔是一个等腰梯形,其形状随月经周期的阶段而变化。从宫底到宫腔变窄至12毫米处的距离为有效长度。这一长度以及子宫腔的总长度可用Wing Sound测量。如果宫内节育器的长度超过有效长度或子宫腔的宽度,其并发症会更频繁。宫内节育器的设计必须适合子宫腔,并符合该器官形状的生理变化。

相似文献

1
Uterine geometry and IUCD design.子宫形态与宫内节育器设计
Br J Obstet Gynaecol. 1982 Sep;89(Suppl 4):1-10. doi: 10.1111/j.1471-0528.1982.tb15064.x.
7
The IUCD in nulliparous women: with special reference to the Multiload copper 250 'mini' and 'short' IUCDs.
Br J Obstet Gynaecol. 1982 Sep;89(Suppl 4):54-7. doi: 10.1111/j.1471-0528.1982.tb15072.x.
8
Haemostasis in menstrual endometrium in the presence of an intrauterine device.
Br J Obstet Gynaecol. 1981 Aug;88(8):825-37. doi: 10.1111/j.1471-0528.1981.tb01310.x.

引用本文的文献

1
Bacterial colonization of intrauterine devices (IUDs).宫内节育器的细菌定植
Arch Gynecol. 1986;239(1):31-7. doi: 10.1007/BF02134286.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。