Suppr超能文献

[一种新型动脉压测量电子设备的研发与临床评估。数字脉搏血压计4000]

[Development and clinical evaluation of a new electronic device for measuring arterial pressure. Sphygmodigital 4000].

作者信息

Baeriswyl C, Muller P O, Regamey C

出版信息

Schweiz Med Wochenschr. 1982 Dec 11;112(50):1829-30.

PMID:7156965
Abstract

Sphygmodigital 4000 is a new automatic electronic device for measurement of the blood pressure (BP). The authors have tested it in their clinic for 6 months and compared the BP values obtained with those measured by the auscultatory method and intraarterially. The differences between the Sphygmodigital 4000 values and those measured by auscultation were -3.7 mm Hg for systolic BP (this error could be eliminated) and +0.2 mm Hg for diastolic BP. The results of comparison with intraarterially measured BPs were -1 mm Hg for systolic BP and +1 mm Hg for diastolic BP. This device is precise and also allows recording of beats.

摘要

数字脉搏血压计4000是一种新型的自动电子血压测量设备。作者在其诊所对该设备进行了6个月的测试,并将其测得的血压值与通过听诊法和动脉内测量法测得的血压值进行了比较。数字脉搏血压计4000测得的收缩压与听诊法测得的收缩压之间的差异为-3.7毫米汞柱(此误差可消除),舒张压之间的差异为+0.2毫米汞柱。与动脉内测量血压结果的比较显示,收缩压差异为-1毫米汞柱,舒张压差异为+1毫米汞柱。该设备精确,还能记录脉搏。