Yatscoff R W, Braidwood J L
Clin Biochem. 1982 Dec;15(6):302-7. doi: 10.1016/s0009-9120(82)96870-9.
A comparison of two column chromatographic (BioRad & Helena), an electrophoretic (Corning) and a colorimetric method for the determination of glycosylated hemoglobin (HbA1) was performed. Blood samples were obtained from a diverse diabetic population and normal persons. Inter-and intra-run precision studies with normal and above normal HbA1 samples indicated that there was no statistically significant difference (p greater than 0.05) in precision among the methods. The electrophoretic and BioRad column chromatographic methods showed best correlation (r = 0.96 p less than 0.001 y = 0.465 + 1.036 x), while the colorimetric method gave HbA1 levels consistently higher than those obtained by electrophoresis (r = 0.852 p less than 0.001 y = 4.99 + 0.867 x). The effects of preincubation to remove labile glycosylated hemoglobin with saline or semicarbazide - aniline preincubation were also studied. There was no statistically significant difference (p greater than 0.05) in the HbA1 levels from erythrocytes after saline preincubation as compared to nontreated samples. Further investigation concerning the value of preincubation to remove the labile HbA1 fraction is warranted.
对两种柱色谱法(伯乐公司和海伦娜公司)、一种电泳法(康宁公司)和一种比色法测定糖化血红蛋白(HbA1)进行了比较。从不同的糖尿病患者群体和正常人中采集血样。对正常和高于正常HbA1水平的样本进行批间和批内精密度研究,结果表明各方法之间在精密度上无统计学显著差异(p大于0.05)。电泳法和伯乐柱色谱法显示出最佳相关性(r = 0.96,p小于0.001,y = 0.465 + 1.036x),而比色法得出的HbA1水平始终高于电泳法测得的水平(r = 0.852,p小于0.001,y = 4.99 + 0.867x)。还研究了用生理盐水预孵育以去除不稳定糖化血红蛋白或用氨基脲 - 苯胺预孵育的效果。与未处理的样本相比,生理盐水预孵育后红细胞中的HbA1水平无统计学显著差异(p大于0.05)。有必要进一步研究预孵育以去除不稳定HbA1组分的价值。