Hawkins D B, Haskell G B
J Speech Hear Disord. 1982 Feb;47(1):71-6. doi: 10.1044/jshd.4701.71.
The amount by which individual functional gain differed from 2 cm3 coupler gain for a single wide-band hearing aid was determined empirically for 20 hearing-impaired adults. Both occluding and non-occluding earmold fittings were evaluated. The amount of this difference accounted for by known measurement differences, such as coupler, real-ear response differences, earmold differences, and sound field pressure differences between the eardrum and hearing aid microphone location, was also established. While it was found that accounting for the known major measurement differences resulted in a reasonable estimate of median differences between measurements of functional gain and measurements of gain in a 2 cm3 coupler, individual variability was quite high.
针对20名听力受损成年人,通过实验确定了单个宽带助听器的个体功能增益与2立方厘米耦合器增益之间的差异量。对封闭和非封闭耳模配件均进行了评估。还确定了由已知测量差异(如耦合器、真耳响应差异、耳模差异以及鼓膜与助听器麦克风位置之间的声场压力差异)所导致的这种差异量。虽然发现考虑已知的主要测量差异能合理估计功能增益测量值与2立方厘米耦合器增益测量值之间的中位数差异,但个体差异相当大。