Schüssler B, Bödeker J, Wallner F, Nagel R, Schmidt-Gollwitzer M
Geburtshilfe Frauenheilkd. 1980 Mar;40(3):233-6. doi: 10.1055/s-2008-1037003.
The measurement of urethral closure pressure with stress is limited by the compressibility of gasiform mediums, which causes damping of intraurethral pressure spikes. In this study damping could be reduced by alterations of the catheter and the connecting tubes of an aircystometer. With this modified technique 55 incontinent women were examined. The differentiation between stress- and urgeincontinence was clinically sufficient.
压力性尿道闭合压的测量受气态介质可压缩性的限制,这会导致尿道内压力峰值出现衰减。在本研究中,通过改变气囊测压计的导管和连接管可减少衰减。采用这种改良技术对55名尿失禁女性进行了检查。在临床上足以区分压力性尿失禁和急迫性尿失禁。