Lund G F, Westbrook R M, Fryer T B
Biotelem Patient Monit. 1980;7(3-4):137-7.
A long-range and long-life telemetry system for heart rate and multiple body temperatures from free-ranging animals is described. This system includes an implantable transmitter, external receiver-retransmitter collar, and a signal conditioner interface circuit to assist in demodulation of receiver tone outputs before data processing. The size of the implant is suitable for animals with body weights of a few kilograms or more; further size reduction of the implant is possible. The ECG is sensed by electrodes designed for internal telemetry and to reduce movement artifacts. The R-wave characteristics are then specifically selected to trigger a short radio frequency pulse. Temperatures are sensed at desired locations by thermistors and then, based on a heart-beat counter, transmitted intermittently via pulse interval modulation. This modulation scheme includes first and last calibration intervals for a reference by ratios with the temperature intervals for as much as a range of 0-50 degrees C with a resolution of 0.1 degrees C. Pulse duration and pulse sequencing are used to discriminate between heart rate and temperature pulses as well as radio frequency interference. The implanted transmitter might be used alone for experiments on animals that frequent particular locations within a large territory, on animals in virtually any laboratory situation, or on animals in moderate-sized enclosures, such as those in a zoological garden. The implanted transmitter is otherwise interfaced with the receiver-retransmitter collar that employs commercial tracking equipment to achieve the long-range transmission. The objective of the design was to achieve a high degree of experimental flexibility and overall high quality in performance. The system was tested in prototype form on a dog.
本文描述了一种用于自由放养动物心率和多个体温的远程、长寿命遥测系统。该系统包括一个可植入式发射器、外部接收 - 转发器项圈以及一个信号调理接口电路,用于在数据处理之前协助解调接收器音频输出。植入物的尺寸适用于体重几千克或更重的动物;进一步减小植入物尺寸是可行的。心电图由专为内部遥测设计的电极感测,并用于减少运动伪影。然后专门选择R波特征来触发短射频脉冲。温度由热敏电阻在所需位置感测,然后基于心跳计数器,通过脉冲间隔调制间歇性地传输。这种调制方案包括用于校准的第一个和最后一个间隔,通过与0 - 50摄氏度范围内的温度间隔的比率进行参考,分辨率为0.1摄氏度。脉冲持续时间和脉冲序列用于区分心率和温度脉冲以及射频干扰。植入式发射器可单独用于对在大片区域内特定位置活动的动物、几乎任何实验室环境中的动物或中等规模围栏(如动物园中的围栏)中的动物进行实验。否则,植入式发射器与采用商业跟踪设备的接收 - 转发器项圈接口,以实现远程传输。该设计的目标是实现高度的实验灵活性和整体高性能。该系统以原型形式在一只狗身上进行了测试。