Sullivan N M, Mayhew J, DiTullio D, Tally F P
J Clin Microbiol. 1978 Oct;8(4):369-73. doi: 10.1128/jcm.8.4.369-373.1978.
Two detection systems for gas-liquid chromatography were compared for the identification of metabolic end products (short-chain organic acids) from anaerobic bacteria. Argon and flame ionization detectors were connected in series with inert argon as the carrier gas for analysis of 35 stock strains and 148 clinical isolates. There was an excellent correlation between the argon and flame ionization chromatograms, and both were readily comparable to the thermal conductivity detector tracings published in the Virginia Polytechnic Institute manual. The linear response of the argon detector was examined by analyzing twofold serial dilutions of the short-chain organic acids and comparing the results to the very linear flame ionization detector. The argon detector was found to react with sufficient linearity within the organic acid concentration range normally obtained from anaerobic bacterial broth cultures. The argon detector, therefore, appears to be a viable alternative to the flame ionization and thermal conductivity detector systems in the identification of anaerobic organisms.
比较了两种用于气相色谱的检测系统,以鉴定厌氧细菌的代谢终产物(短链有机酸)。氩气和火焰离子化检测器串联连接,以惰性氩气作为载气,用于分析35株标准菌株和148株临床分离株。氩气色谱图和火焰离子化色谱图之间具有极好的相关性,并且两者都很容易与弗吉尼亚理工学院手册中公布的热导检测器图谱进行比较。通过分析短链有机酸的两倍系列稀释液并将结果与线性非常好的火焰离子化检测器进行比较,研究了氩气检测器的线性响应。发现在通常从厌氧细菌肉汤培养物中获得的有机酸浓度范围内,氩气检测器具有足够的线性反应。因此,在鉴定厌氧生物方面,氩气检测器似乎是火焰离子化和热导检测器系统的可行替代方案。