Hull C J, Sibbald A
Br J Anaesth. 1981 Apr;53(4):385-91. doi: 10.1093/bja/53.4.385.
A demand analgesia apparatus is described. It communicates with the patient by means of spoken messages in any desired language, so that preliminary instruction is unnecessary. Dosage is limited by a reduction in respiratory rate, and by a series of electronic fail-safe circuits. Operation of the system is illustrated in a series of 10 patients given fentanyl "on demand" for the first 12 h after operation. The mean dose rate 0.73 +/- 0.49 microgram min-1. There was no evidence of cumulation or tolerance during the period of study.
本文描述了一种按需镇痛装置。它能通过任何所需语言的语音信息与患者进行沟通,因此无需预先指导。剂量通过呼吸频率的降低以及一系列电子故障安全电路来限制。该系统的操作在10例患者中得到了展示,这些患者在术后的前12小时按需使用芬太尼。平均剂量率为0.73±0.49微克/分钟。在研究期间没有累积或耐受的迹象。