Sankaran A, Gokarn R S, Gangadharan P
Br J Radiol. 1981 Apr;54(640):328-34. doi: 10.1259/0007-1285-54-640-328.
This instrument was developed in response to a requirement for an accurate, stable and portable transfer dosemeter for calibration, at therapy dose levels, of equipment used for generating X and gamma rays, electrons and neutrons. The detector is a 0.5 cm3 ionization chamber capable of fitting various wall materials reproducibly at the end of the chamber stem. The measuring system uniquely combines the features of a MOSFET electrometer and an automatic Townsend balance. When used for X, gamma and neutron radiations, the instrument measures the tissue kerma in free air on two ranges: 0.001 - 1.999 Gy (0.1 - 199.9 rad) and 0.01 - 19.99 Gy (1 - 1999 rad) or their exposure equivalents, with autoranging feature when the first range is exceeded. The polarizing voltage (180 V) can be reversed for electron and neutron dosimetry. The dosemeter has a measuring accuracy of +/- 0.2% FS +/- 1 digit and operates on four 1.5 V torchlight cells or on AC mains (200-250 V, 50 - 60 Hz). It utilizes solid state devices, CMOS integrated circuits and displays, and is not affected by RF fields. The instrument is enclosed in a brief-case for portability and is easy to operate and maintain in a hospital.
本仪器是应一项需求而开发的,该需求是要有一种准确、稳定且便携的传输剂量仪,用于在治疗剂量水平下校准产生X射线、γ射线、电子和中子的设备。探测器是一个0.5立方厘米的电离室,能够在电离室杆的末端可重复地安装各种壁材料。测量系统独特地结合了MOSFET静电计和自动汤森平衡的特性。当用于X射线、γ射线和中子辐射时,该仪器在两个量程上测量空气中的组织比释动能:0.001 - 1.999 Gy(0.1 - 199.9拉德)和0.01 - 19.99 Gy(1 - 1999拉德)或其照射当量,当超过第一个量程时具有自动量程功能。极化电压(180 V)可反转,用于电子和中子剂量测定。剂量仪的测量精度为±0.2%满量程±1位数字,由四个1.5 V手电筒电池或交流市电(200 - 250 V,50 - 60 Hz)供电。它采用固态器件、CMOS集成电路和显示器,不受射频场影响。仪器装在一个公文包内以便于携带,在医院中易于操作和维护。