We evaluated the performance of a CK isoenzyme kit utilizing agarose electrophoresis with fluorescence scanning of an overlay paper rather than the gel. We compared this method to ion exchange chromatography with mini-columns of DEAE Sephadex using samples from hospitalized patients and from a reference population of ambulatory outpatients with common, chronic medical conditions. The detection limit, precision and linearity of the electrophoretic method were acceptable, but the method overestimated CK-MB, requiring a different reference range for clinical application. There were inter-racial differences within the reference population for total CK and for the percent of CK-MB using either method. We discuss the significance of our findings with respect to clinical decision limits for CK-MB.

摘要

我们评估了一种肌酸激酶同工酶试剂盒的性能,该试剂盒采用琼脂糖电泳,并对覆盖纸而非凝胶进行荧光扫描。我们将此方法与使用DEAE葡聚糖微型柱的离子交换色谱法进行比较,使用的样本来自住院患者以及患有常见慢性疾病的非住院门诊参考人群。电泳法的检测限、精密度和线性度是可接受的,但该方法高估了肌酸激酶MB型(CK-MB),临床应用时需要不同的参考范围。在参考人群中,无论使用哪种方法,总肌酸激酶(total CK)以及CK-MB百分比均存在种族差异。我们讨论了研究结果对于CK-MB临床决策限值的意义。