[尿液标本的自动直接抗菌药敏试验]

[Automated direct antimicrobial susceptibility testing of urine specimens].

作者信息

Magliani W, Goffi M, Conti S, Toni M

出版信息

Quad Sclavo Diagn. 1980 Dec;16(4):428-36.

PMID:7244113
Abstract

Two microscopical screening methods have been used to assay the suitability of urine samples for automated direct antimicrobial susceptibility testing by the means of a "MS-2 Microbiology System" apparatus. Gram stained smears showed if specimens were monomicrobic and fresh-mount bacterioscopy gave information about presumptive counts to establish inoculum size. The results obtained by the automated method have been compared with those of the direct disk diffusion method and the classical Stokes method on a single isolated colony. The three systems were reliable and only 7% of specimens gave major discrepancies and 15% minor errors. The influence of each antibiotic drug and of each bacterial species on the discrepancies has been discussed. Direct susceptibility testing by MS-2 seems to be accurate and in the same time so fast (5-6 h vs. 18-24 h) to appear very useful in the rational guide of chemotherapy in the more severe UTI.

摘要

已采用两种显微镜筛选方法,通过“MS-2微生物系统”仪器来测定尿液样本是否适合进行自动直接药敏试验。革兰氏染色涂片可显示标本是否为单一微生物感染,而新鲜涂片细菌镜检则可提供有关推测菌数的信息,以确定接种量。已将自动方法获得的结果与直接纸片扩散法和经典斯托克斯法对单个分离菌落的结果进行了比较。这三种系统都很可靠,只有7%的标本存在重大差异,15%存在轻微误差。已讨论了每种抗生素药物和每种细菌种类对差异的影响。MS-2直接药敏试验似乎准确,同时又非常快速(5-6小时对比18-24小时),在更严重的尿路感染化疗合理指导方面显得非常有用。