Moyer R F
Radiology. 1981 Jul;140(1):237-9. doi: 10.1148/radiology.140.1.7244231.
An aluminum wedge was designed to facilitate evaluation of the constancy of high-energy electron teletherapy beams (6-18 MeV). Ready-packed film was exposed with the wedge on top, and optical density distribution was recorded with a scanning densitometer. Energy-dependent statistics were obtained using the practical-range extrapolation method. The procedure takes approximately 15 minutes and requires less than one minute of therapy machine time. Repeated daily measurements on a medical linear accelerator at 12 MeV showed a variation of +/- 0.37 MeV (3 sigma). This technique has proved useful for prompt machine evaluation following malfunctions and has been incorporated into routine quality assurance testing.
设计了一种铝楔形物,以方便评估高能电子远距离治疗束(6 - 18兆电子伏特)的稳定性。将预包装胶片放在楔形物上方进行照射,并用扫描密度计记录光密度分布。使用实际射程外推法获得与能量相关的统计数据。该程序大约需要15分钟,且治疗机时间不到1分钟。在一台12兆电子伏特的医用直线加速器上每天重复测量显示,变化为±0.37兆电子伏特(3西格玛)。事实证明,该技术对于故障后快速进行机器评估很有用,并已纳入常规质量保证测试中。