Suppr超能文献

使用铝楔形物对电子束能量进行质量保证。

Quality assurance of electron-beam energy using an aluminum wedge.

作者信息

Moyer R F

出版信息

Radiology. 1981 Jul;140(1):237-9. doi: 10.1148/radiology.140.1.7244231.

Abstract

An aluminum wedge was designed to facilitate evaluation of the constancy of high-energy electron teletherapy beams (6-18 MeV). Ready-packed film was exposed with the wedge on top, and optical density distribution was recorded with a scanning densitometer. Energy-dependent statistics were obtained using the practical-range extrapolation method. The procedure takes approximately 15 minutes and requires less than one minute of therapy machine time. Repeated daily measurements on a medical linear accelerator at 12 MeV showed a variation of +/- 0.37 MeV (3 sigma). This technique has proved useful for prompt machine evaluation following malfunctions and has been incorporated into routine quality assurance testing.

摘要

设计了一种铝楔形物,以方便评估高能电子远距离治疗束(6 - 18兆电子伏特)的稳定性。将预包装胶片放在楔形物上方进行照射,并用扫描密度计记录光密度分布。使用实际射程外推法获得与能量相关的统计数据。该程序大约需要15分钟,且治疗机时间不到1分钟。在一台12兆电子伏特的医用直线加速器上每天重复测量显示,变化为±0.37兆电子伏特(3西格玛)。事实证明,该技术对于故障后快速进行机器评估很有用,并已纳入常规质量保证测试中。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验