Myers J L, Bull A, Kastl D G, Pierce W S
J Thorac Cardiovasc Surg. 1981 Aug;82(2):263-7.
We recently utilized a left ventricular assist device in a patient who could not be separated from cardiopulmonary bypass after aortic and mitral valve replacement. The nonfunctioning pericardial aortic valve became completely occluded owing to a fibrinous fusion of the valve leaflets during the 8-day period of assist pumping. A second operation was required to open the fused valve leaflets. This experience suggests that special care should be taken in selecting an aortic valve prosthesis in a patient who may require the left ventricular assist device in order to be separated from cardiopulmonary bypass.
我们最近对一名在主动脉瓣和二尖瓣置换术后无法脱离体外循环的患者使用了左心室辅助装置。在辅助泵血的8天期间,无功能的心包主动脉瓣因瓣叶的纤维蛋白融合而完全闭塞。需要进行第二次手术来打开融合的瓣叶。这一经验表明,对于可能需要左心室辅助装置以脱离体外循环的患者,在选择主动脉瓣假体时应格外小心。