Jester J V, Venet T, Lee J, Schanzlin D J, Smith R E
Am J Ophthalmol. 1981 Aug;92(2):172-7. doi: 10.1016/0002-9394(81)90765-0.
We performed radial keratotomy using the Fyodorov techniques on 14 Eye Bank eyes. Using stepwise regression and best subset statistical analysis, we examined the effects of incision length. We found that the most significant variables affecting postoperative corneal curvature were the incision depth and length. The histologic examination of serial sections of the Eye Banks eyes after this procedure revealed that the average achieved incision depth, using a standard blade depth setting, had variations of up to 30% between eyes. This inability to standardize incision depth may lead to the variable surgical results reported in radial keratotomy.