Abstract

We describe a fully automated method of analysis for plasma epinephrine and norepinephrine by use of a two-column system of "high-performance" liquid chromatography. Catecholamines in deproteinized plasma are purified on the first (preparation) column, then transferred automatically to the second (analytical) column in which epinephrine and norepinephrine are resolved. These compounds are then determined fluorometrically with a continuous-flow reaction system by the trihydroxyindole method. The minimum sensitivity of the method is 0.02 and 0.04 pmol for epinephrine and norepinephrine, respectively. One assay can be completed in 30 min, and greater than 1 mL of plasma is required for the procedure. The within-run CV in the chromatographic determination of pooled plasma was greater than 3%, and the analytical overall recovery of the two compounds was 95%. The concentrations in plasma of medical students at rest were 0.32 (SE 0.06) nmol/L for epinephrine and 0.98 (SE 0.08) nmol/L for norepinephrine.

摘要

我们描述了一种使用“高效”液相色谱双柱系统对血浆肾上腺素和去甲肾上腺素进行全自动分析的方法。去蛋白血浆中的儿茶酚胺在第一根(制备)柱上进行纯化,然后自动转移至第二根(分析)柱,在该柱上分离肾上腺素和去甲肾上腺素。然后通过三羟基吲哚法,利用连续流动反应系统对这些化合物进行荧光测定。该方法对肾上腺素和去甲肾上腺素的最低检测灵敏度分别为0.02和0.04皮摩尔。一次检测可在30分钟内完成,该过程需要超过1毫升的血浆。在对混合血浆进行色谱测定时,批内变异系数大于3%,两种化合物的分析总回收率为95%。医学生静息时血浆中肾上腺素的浓度为0.32(标准误0.06)纳摩尔/升,去甲肾上腺素的浓度为0.98(标准误0.08)纳摩尔/升。