Dito W R, Tucker E S, Nakamura R M
Am J Clin Pathol. 1981 Dec;76(6):753-9. doi: 10.1093/ajcp/76.6.753.
A comparative evaluation of a laser nephelometer (J.T. Baker Immunology Series 420 Laser Nephelometer) with other immunoprecipitin methods in each of two separate laboratories is reported. This is a batch-oriented almost fully-automated laser nephelometer capable of both kinetic and end-point nephelometry. In the kinetic mode, it will automatically detect antigen-excess and perform multiple reassays of further dilutions of sample in a completely unattended mode. Both instrument maintenance and down-time were minimal in both laboratories. The data derived from manufacturer's kits for the assay of IgG, IgA, and IgM were compared to both commercial radial-immunodiffusion and a two-point fixed interval immunoturbidimetric method on a centrifugal fast analyzer. Correlation coefficients exceeded 0.9 in all instances, with variable bias depending on the assay. The apparent bias was felt to be due to a combination of factors including different antibody sources, variance in reference material, assay conditions, difference in measurement modality, and form of mathematical evaluation. The instrument was easy to operate, rapidly gained technologist acceptance, replicated various assay levels at coefficient of variations of less than 10% (between run) and fulfilled our expectations regarding the automatic detection of antigen excess and unattended sample reassay.
本文报告了在两个独立实验室中,对激光散射比浊仪(J.T.贝克免疫系列420激光散射比浊仪)与其他免疫沉淀方法进行的比较评估。这是一款面向批次的几乎全自动的激光散射比浊仪,能够进行动力学和终点比浊法检测。在动力学模式下,它将自动检测抗原过量情况,并在完全无人值守的模式下对样品的进一步稀释进行多次重新检测。在两个实验室中,仪器维护和停机时间都很少。将制造商试剂盒用于检测IgG、IgA和IgM的数据与商业放射免疫扩散法以及离心快速分析仪上的两点固定间隔免疫比浊法进行了比较。在所有情况下,相关系数均超过0.9,偏差因检测方法而异。明显的偏差被认为是由多种因素共同作用导致的,包括不同的抗体来源、参考物质的差异、检测条件、测量方式的不同以及数学评估形式。该仪器操作简便,迅速获得了技术人员的认可,在批间变异系数小于10%的情况下能够重复各种检测水平,并且满足了我们对抗原过量自动检测和无人值守样品重新检测的期望。