Combie J, Blake J W, Ramey B E, Tobin T
Am J Vet Res. 1981 Sep;42(9):1523-30.
Morphine was detected in equine biological fluids by a combination of liquid-liquid extraction and column chromatography, followed by derivatization and gas-liquid chromatographic assay, using electron capture detector. Recovery of morphine from the equine biological samples was poor. However, despite an overall recovery of less than 20%, this method had a detection limit of 0.2 ng/ml. Addition of 5,000 U of bovine liver beta-glucuronidase/ml of urine enabled detection of the drug in urine for up to 144 hours after horses were given 0.1 mg of morphine/kg of body weight. Morphine was found for at least 24 hours in serum samples. An adaptation of logit-log transformation of gas-liquid chromatographic data for linearization over 3 log units suggested a simple adaptation to existing semiautomated data handling systems.
通过液液萃取和柱色谱法相结合,继以衍生化和使用电子捕获检测器的气液色谱测定法,在马的生物体液中检测到吗啡。从马的生物样品中回收吗啡的效果较差。然而,尽管总体回收率低于20%,该方法的检测限为0.2纳克/毫升。每毫升尿液中添加5000单位的牛肝β-葡萄糖醛酸酶,能够在给马按每千克体重0.1毫克的剂量注射吗啡后长达144小时的尿液中检测到该药物。在血清样品中至少24小时都能检测到吗啡。对气液色谱数据进行logit-log变换以在3个对数单位上实现线性化的一种方法,表明对现有的半自动数据处理系统进行简单改造即可。