Gressner A M, Wallraff P
J Clin Chem Clin Biochem. 1981 Dec;19(12):1213-5.
A commercially available test kit for the kinetic immunoturbidimetric measurement of cold insoluble globulin was evaluated. The procedure is precise (CV intra-assay between 2.7 and 5.1%, CV inter-series 3.8%), simple, rapid, and can be fully mechanized. Antigen concentrations determined by both turbidimetry and laser immunonephelometry correlate with r= 0.94 (y = 0.96 x + 40.64). The detection limit is about 50 mg/1 fibronectin which is nearly 40 times higher than that of laser nephelometry. Improvements of the turbidimetric procedure are achieved by reduction of sample and reaction volume, higher predilution of antiserum, and by use of the peak rate modification (instead of fixed time) of the kinetic test. The peak rate version allows the determination of fibronectin in plasma within 3 to 5 min.
对一种市售的用于冷不溶性球蛋白动力学免疫比浊法测量的检测试剂盒进行了评估。该方法精确(批内变异系数在2.7%至5.1%之间,系列间变异系数为3.8%)、简单、快速,并且可以完全机械化。通过比浊法和激光免疫比浊法测定的抗原浓度具有相关性,r = 0.94(y = 0.96x + 40.64)。检测限约为50mg/1纤维连接蛋白,这比激光比浊法高出近40倍。通过减少样品和反应体积、提高抗血清的预稀释度以及采用动力学检测的峰值速率修正法(而非固定时间法)对比浊法进行了改进。峰值速率法可在3至5分钟内测定血浆中的纤维连接蛋白。