Marinoni G P, Pasquale C, Curti M T, Assandri J, Teodoro C, Collarini L, De Filippi R
G Ital Cardiol. 1981;11(12):1982-7.
Four cases of subjects with implanted pacemakers are reported. The patients temporarily presented suspected symptoms of altered activity of the pacemakers. On the other hand usual tests demonstrated a regular activity of the pacemakers. The dynamic electrocardiography, by a regular registration during 24/48 hours and during usual working, documented; once a partial hole in the circuits of the generator, twice a partial break of the catheter, once a displacement. It is underlined the importance of dynamic electrocardiography to identify temporarily altered activity of pacemakers, undiagnosticable by disposable usual tests.