Roesink H D, Koenhen D M, DeJongh M A, Smolders C A
Ultramicroscopy. 1978;3(3):325-8. doi: 10.1016/s0304-3991(78)80042-4.
In this study the cryo-unit is introduced as a new and useful instrument to investigate water-containing specimens with the SEM. Often water-containing biological specimens are studied, but in our case we used water-swollen polymer membranes. The results show that application of a cryo-unit permits the study of this material at low temperatures up to magnifications of about 10 000 times, while other techniques failed to give reproducible results.
在本研究中,低温装置被作为一种新型且实用的仪器引入,用于通过扫描电子显微镜研究含水标本。人们经常对含水生物标本进行研究,但在我们的案例中,我们使用的是水溶胀聚合物膜。结果表明,使用低温装置能够在低温下对这种材料进行研究,放大倍数可达约10000倍,而其他技术则无法给出可重复的结果。