Suppr超能文献

人工晶状体植入术后无菌性前房积脓

Sterile hypopyon following intraocular lens surgery.

作者信息

Meltzer D W

出版信息

Arch Ophthalmol. 1980 Jan;98(1):100-4. doi: 10.1001/archopht.1980.01020030102008.

Abstract

Energy-dispersive x-ray analysis (EDXA) was used to study nine intraocular lenses from one manufacturer. Residual polishing compound was found on two of three dry-sterilized lenses from a "hot lot" associated with a high incidence of sterile hypopyon. Similar material was not present on normal lenses. This polishing compound was apparently fused to the lens surface, preventing its removal by ultrasonic cleaning. In addition, three wet-sterilized lenses were found to be covered with small crystals containing barium. These crystals apparently resulted from leaching of the glass storage vial by the caustic storage fluid. It is suggested that some cases of sterile hypopyon following intraocular lens surgery may be caused by residual polishing compound present on the lens surface.

摘要

能量色散X射线分析(EDXA)被用于研究来自一家制造商的9个人工晶状体。在来自一个与无菌性前房积脓高发生率相关的“热批次”的三个干式灭菌晶状体中的两个上发现了残留的抛光化合物。正常晶状体上不存在类似物质。这种抛光化合物显然与晶状体表面融合,阻止了通过超声清洗将其去除。此外,发现三个湿式灭菌晶状体覆盖有含钡的小晶体。这些晶体显然是由腐蚀性储存液对玻璃储存瓶的浸出导致的。有人提出,人工晶状体手术后的一些无菌性前房积脓病例可能是由晶状体表面存在的残留抛光化合物引起的。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验