Sterile hypopyon secondary to ultrasonic cleaning solution.

作者信息

Richburg F A, Reidy J J, Apple D J, Olson R J

出版信息

J Cataract Refract Surg. 1986 May;12(3):248-51. doi: 10.1016/s0886-3350(86)80002-5.

Abstract

During a 20-week period, 21 cases of sterile hypopyon occurred 24 hours after uncomplicated extracapsular cataract extraction and in-the-bag posterior chamber intraocular lens implantation at the Valley Eye Institute. While all cases responded to steroid treatment and exhibited no residual ocular effects, the cause of the hypopyon remained a mystery despite an on-going reevaluation of the surgical procedure and sterilization techniques. Several suspected causes were investigated and eliminated. It was clinically determined that surgical instruments placed in older ultrasonic cleaning solution caused the hypopyons. Further analysis at the Center for Intraocular Research at the University of Utah revealed that the instruments were contaminated by a heat-stable endotoxin that remained on the instruments through the autoclaving process. Surgeons currently using ultrasonic instrument cleaners should be aware that, with some models, following the manufacturer's recommended fluid changing procedures could result in endotoxic contamination of surgical instruments which could, in turn, cause severe inflammatory reactions in the immediate postoperative period.

摘要

在20周的时间里,山谷眼科研究所21例无菌性前房积脓病例发生在无并发症的囊外白内障摘除及后房型人工晶状体植入术后24小时。尽管所有病例对类固醇治疗均有反应且未出现残留眼部影响,但尽管对手术过程和消毒技术进行了持续重新评估,前房积脓的原因仍是个谜。对几个可疑原因进行了调查并排除。临床确定,放置在旧的超声清洗液中的手术器械导致了前房积脓。犹他大学眼内研究中心的进一步分析表明,器械被一种热稳定内毒素污染,这种内毒素在高压灭菌过程中仍残留在器械上。目前使用超声器械清洗器的外科医生应意识到,对于某些型号,遵循制造商推荐的换液程序可能会导致手术器械的内毒素污染,进而可能在术后即刻引起严重的炎症反应。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验