Alderman F R, Gitelman H J
Clin Chem. 1980 Feb;26(2):258-60.
We describe an improved method for determination of aluminum in serum by atomic absorption spectrophotometry with a graphite furnace. Interferences have been controlled and total combustion of the samples achieved by the use of new diluent in a molybdenum-treated graphite tube. The detection limit of this method is 2 microgram/L.